User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HP Model E2377A Verification Procedure DATE: 18-Aug-92 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 99% NUMBER OF TESTS: 32 NUMBER OF LINES: 92 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ B 1.002 ASK- R J W 1.003 MMFC C2 1.004 HEAD { Initial Process } 1.005 DISP ! Plug-in the model E2377A and turn it on. 1.006 DISP ! Before performing the Verification test 1.006 DISP ! allow the unit time to warm-up. 1.007 HEAD { Zero Test } 1.008 DISP Place shorting strap across input to UUT 1.009 DISP Set the UUT to Autorange VDC range 1.010 EVAL Does the {UUT read ZERO ñ2 Digits ?} 2.001 DISP Set the UUT to Autorange Ohms range 2.002 EVAL Does the {UUT read ZERO ñ2 Digits ?} 3.001 HEAD 3.002 DISP [27][91]1m Connect as follows 3.002 DISP ! ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ 3.002 DISP ! [27][91]1m 5700 UUT 3.002 DISP ! Sense HI ÄÄÄÄÄÄÄÄÄ¿ 3.002 DISP ! Output HI ÄÄÄÄÄÄÄÄÄÁÄ> Volt/Ohm Input 3.002 DISP ! Output LO ÄÄÄÄÄÄÄÄÄÂÄ> Common 3.002 DISP ! Sense LO ÄÄÄÄÄÄÄÄÄÙ 3.002 DISP ! AUX Out ÄÄÄÄÄÄÄÄÄÄÄ> UUT HIGH INPUT 3.003 HEAD { DC Voltage Test } 3.004 DISP Select the Volts DC function 3.005 5700 300mV 0.3% 2W 4.001 5700 3.00V 0.3% 2W 5.001 5700 30.0V 0.4% 2W 6.001 5700 300V 0.4% 2W 7.001 5700 1000V 0.4% 2W 8.001 HEAD { AC Voltage test } 8.002 DISP Select the Volts AC function 8.003 5700 3.00V 1% 100H 2W 9.001 5700 3.00V 1% 600H 2W 10.001 5700 30.0V 1% 100H 2W 11.001 5700 30.0V 2% 600H 2W 12.001 5700 300V 1% 100H 2W 13.001 5700 300V 2% 600H 2W 14.001 5700 750V 1% 100H 2W 15.001 5700 750V 2% 600H 2W 16.001 HEAD { Ohms Test } 16.002 DISP Select the Ohms function 16.003 5700 190Z 0.7% RW 17.001 5700 1.9KZ 0.7% RW 18.001 5700 19KZ 0.7% RW 19.001 5700 190KZ 0.7% 2W 20.001 5700 1.9MZ 0.7% 2W 21.001 5700 19MZ 2% 2W 22.001 HEAD { DC Current test } 22.002 DISP Select the DC uA function 22.003 5700 300uA 1% 2W 23.001 DISP Select the DC mA function 23.002 5700 3mA 1% 2W 24.001 5700 30mA 1% 2W 25.001 5700 300mA 1.5% 2W 26.001 HEAD { AC Current test } 26.002 DISP Select the AC uA function 26.003 5700 300uA 2% 100H 2W 27.001 DISP Select the AC mA function 27.002 5700 3mA 2% 100H 2W 28.001 5700 30mA 2% 100H 2W 29.001 5700 300mA 2% 100H 2W 30.001 HEAD 30.002 DISP DISCONNECT AUX CURRENT OUT FROM THE UUT 30.003 DISP 30.004 DISP ![27][91]1m Connect High Current as follows 30.004 DISP ! ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ 30.004 DISP ! 5725A UUT 30.004 DISP ! Output HI ÄÄÄÄÄÄÄÄÄÄÄ> Fused terminal (10A) 30.004 DISP ! Output LO ÄÄÄÄÄÄÄÄÄÄÄ> Common 30.005 HEAD { Boosted DC Current test } 30.006 DISP Select the 10 Amp DC function 30.007 5700 10.0A 1.5% B1 2W 31.001 HEAD { Boosted AC Current test } 31.002 DISP Select the 10 Amp AC function 31.003 5700 10.0A 2% 100H B1 2W 32.001 HEAD 32.002 DISP ! P R O C E D U R E C O M P L E T E 32.002 DISP ! Disconnect all cables 32.002 DISP ! [27][91]5m Complete required paperwork